Gastrointestinal Intervention

A review of recent experience with transjugular intrahepatic portosystemic shunt creation using intravascular ultrasound

Additional article information

Abstract

Portal vein access has historically been the most technically challenging step in the creation of transjugular intrahepatic portosystemic shunts (TIPSs). The use of intravascular ultrasound (IVUS) for guidance of portal vein access during TIPS creation has garnered much interest in recent years. Recent literature has suggested potential improvements in procedural metrics that may result from use of IVUS for TIPS. This review aims to provide historical context, detail technical advances and describe recent clinical experience with the use of IVUS for TIPS creation.

Keywords: Hypertension, potal, Intravascular ultrasound, Transjugular intrahepatic portosystemic shunt

Introduction

Prior to the clinical introduction of transjugular intrahepatic portosystemic shunt (TIPS) in the late 1980s, decompression of the portal system necessitated creation of an extrahepatic surgical shunt. The concept of a TIPS was initially conceived after inadvertent portal vein puncture during transjugular cholangiography in the late 1960s. The first report of successful transjugular portosystemic shunt creation was published in 1969.1 The shunts were composed of various materials such as Teflon, silastic and spring coil tubing coated with silicone-based copolymer. However, shunt patency was a major hindrance to wider adoption as most of the shunts occluded within hours to days after placement. The advent of bare metallic stents heralded the wider clinical availability of TIPS for the treatment of sequelae of portal hypertension in the 1990s. The subsequent development of expandable polytetra-fluoroethylene covered stents further improved shunt patency.2,3 Today, TIPS is a well established therapeutic modality for (1) refractory variceal bleeding and (2) ascites in patients who do not tolerate repeated large volume paracentesis. Indications for TIPS have continued to evolve as improved understanding of liver disease and technical refinements have altered treatment algorithms in certain situations. Evidence for the use of TIPS to treat other conditions, including primary prophylaxis in patients with high risk acute variceal bleeding, gastric and ectopic variceal bleeding, primary treatment of medically refractory ascites, recurrent ascites following liver transplantation, hepatorenal syndrome, Budd-Chiari syndrome and portal vein thrombosis continues to be accumulated.4 TIPS is classically one of the more challenging procedures in interventional radiology and has required relatively high radiation doses and contrast volumes.5 Much of the difficulty lies in manipulating and guiding a puncture needle towards the right portal vein under limited two-dimensional fluoroscopic guidance. Intravascular ultrasound (IVUS) has been described as an adjunctive real-time imaging modality that may potentially facilitate portal vein access.

Portal Venous Access

Conventional transjugular portosystemic shunt creation technique involves obtaining vascular access via the right internal jugular vein, although some operators have found it easier to cannulate the right hepatic vein from the left internal jugular vein.6 Through the jugular access, a liver access set (Rosch-Uchida, Haskal or Ring; Cook Medical, Bloomington, IN, USA) is placed into the right hepatic vein. The Rosch Uchida liver access set contains a 10 Fr 40 cm introducer sheath and an inner 14 gauge 51.5 cm stiffening cannula through which a 5 Fr catheter and an inner 0.038 inch 62.5 cm trocar stylet assembly are placed. The liver access set is then rotated towards the expected location of the right portal vein—usually anterior in the standard anatomical configuration. Once the entire assembly is well-positioned in the right hepatic vein and wedged against the vein wall, the 5 Fr catheter and stylet assembly is advanced towards the right portal vein. Operators often feel a “popping” sensation when they enter the portal vein. The assembly is then advanced 1 cm beyond the portal vein and the stylet removed from the 5 Fr catheter. The catheter is withdrawn and aspirated until blood return is observed. Contrast is injected to confirm catheter position within the portal system.

Due to the limited visualization provided by traditional two dimensional fluoroscopy systems, TIPS may often involve several needle punctures, and up to 35 needle punctures in a single procedure have been reported.7 Hepatic arterial and other non-target organ puncture (Fig. 1) has been described to occur in 1% to 6% of all TIPS cases.8,9 Clinical sequelae of these inadvertent needle punctures generally tend to be limited and are symptomatic in less than 2% of cases.9,10 Occasionally hepatic arterial injury leads to significant morbidity and even mortality.11 In one case report, a patient became hemodynamically unstable four hours following a TIPS procedure that had been complicated by hepatic arterial cannulation and injury. CT and conventional angiography demonstrated arterial extravasation and plug embolization was performed at a right hepatic arterial branch. Despite recovery of hemodynamic stability, the patient’s serum bilirubin continued to rise and the patient eventually died of fulminant liver failure.12 Biliary puncture occurs in fewer than 5% of TIPS cases. However, biliary transection raises the risk of fistula formation and consequently hemobilia, cholangitis, sepsis, stent infection or stent occlusion secondary to pseudointimal hyperplasia.8 Portal venous access is particularly difficult with narrowed portal veins, obstructing tumor mass, unfavorable vascular anatomy, hepatic and/or portal venous thrombosis. Unfavorable vascular anatomy is often related to advanced cirrhosis, whereby the standard positional relationship between the right hepatic vein and right portal vein is lost, but can also be related to narrowed hepatic veins due to Budd-Chiari syndrome or other veno-occlusive disorders.13 Tumor masses between the right hepatic vein and portal vein preclude standard TIPS placement due to the risk of hemorrhage and tumor seeding associated with often highly vascular tumors.

Figure F1
(A) Hepatic arterial cannulation (arrowhead) and associated colonic puncture with resulting intraluminal contrast leakage (arrow) during a conventional transjugular intrahepatic portosystemic shunt procedure. (B) Follow-up computed tomography performed 6 hours ...

In the initial report published in 1969, portal vein cannulation was assisted by balloon catheter placement in the left portal vein via direct vascular access following laparotomy.1 Many techniques have since been described to assist with portal vein cannulation. Indirect portography involves injecting contrast into the superior mesenteric artery and taking delayed portovenous phase images.14 While this technique is technically uncomplicated, drawbacks of this technique include use of large volumes of contrast, need for additional arterial access, and poor quality of images due to contrast dilution and possible patient motion. As an alternative to indirect arterial portography, wedged portography using direct injection of iodinated contrast or CO2 via a wedged cannula in a hepatic vein has been described. While these images tend to be of superior quality, the rate of subsequent contrast extravasation has been reported to range from 1.8% with CO2 to 7.5% with iodinated contrast.15 Contrast extravasation, while clinically insignificant in the majority of cases, may portend life-threatening hepatic capsular rupture.8,9 Placement of marker wires in the right hepatic artery via the celiac axis has been described with mixed success. In one series of 11 patients, one patient suffered from hepatic arterial injury and another from biliary ductal injury due to inadvertent needle puncture.16 In certain cases, the umbilical vein can be cannulated for a contrast portogram. However, this approach necessitates an adequately sized recanalized umbilical vein and an additional venous access site. Some centers have found success with transhepatic or trans-splenic placement of marker wires within the portal veins as a target for needle puncture.17 One of the drawbacks of this technique is the additional needle puncture through highly vascular organs with associated risks of hemorrhage and organ damage. Due to the risks, many operators perform embolization along the tract. With the exception of CO2 wedged portograms, the use of the aforementioned techniques has been generally reserved for special circumstances and TIPS is performed today with similar technique since its clinical introduction in the 1990s.

Intravascular Ultrasound

IVUS systems were first introduced in the 1980s. Initial transducers were designed to evaluate blood vessel walls. As a result, they were high frequency, in the 20 to 40 MHz range, with limited tissue penetration. In the early 2000s, lower frequency 9 MHz transducers were developed to image cardiac chambers during electrophysiology procedures. These systems featured mechanical ultrasound transducer tipped catheters which rotated at 1,800 rpm to provide a cross sectional view in the radial plane with 5 cm tissue penetration (Fig. 2). While it provided a comprehensive view of the anatomy at each level, the 360° image suffered from high levels of anatomical distortion and lack of color and pulse wave Doppler capabilities.18

Figure F2
A 51-year-old female with a history of hepatitis C virus infection status post remote orthotopic liver transplant complicated by recurrent cirrhosis, esophageal varices and presenting with hematochezia and hematemesis. (A–C) ...

More recently, phased array ultrasound transducers with a 90° longitudinal side firing view have been developed. The ACUSON AcuNav system (Siemens Healthineers, Erlangen, Germany) features an 8 or 10 Fr diameter and 90 cm long catheter containing an phased array ultrasound transducer at the distal tip. The transducer has a variable 5 to 10 MHz frequency with up to 15 cm tissue penetration. The ViewFlex Xtra Intracardiac Echocardiography catheter (St. Jude Medical, St. Paul, MN, USA) features a 9 Fr 90 cm ultrasound-tipped catheter with a fixed 9 MHz frequency providing a roughly 18 cm tissue penetration. Both of these catheter probes can be steered at the distal tip in the anterior-posterior and medial-lateral planes and provide functional analysis with color and pulse wave Doppler. When placed at the level of the retrohepatic IVC, the side-firing ultrasound probes provide a longitudinal 90° view of the hepatic parenchyma and vessels. The portal access needle tip can be visualized in real time as the transhepatic access set inner cannula and stylet assembly is advanced from the hepatic vein to the portal vein. Further, there is no need to advance the assembly beyond the portal vein as the needle tip can be visualized in real-time as it enters the portal vein (Fig. 3).

Figure F3
A 57-year-old obese female with hematemesis secondary to extensive acute mesenteric, splenic and portal venous thrombosis following sleeve gastrectomy. (A, B) Computed tomography performed prior to the procedure demonstrated acute ...

Current Literature

The use of IVUS in hepatic interventions was first described for direct portocaval shunting whereby a shunt is placed between the portal vein and the retrohepatic IVC through the caudate lobe. Its use for TIPS guidance was first described in 2004.19 Using the aforementioned AcuNav catheter, the authors were able to place TIPS in three swine after one or two needle punctures without the use of wedged portography. No post-procedural complication was reported. The first retrospective clinical series of 25 cases using IVUS compared to 75 cases using conventional technique was published in 2012.20 Cases were performed using left internal jugular vein for placement of the liver access set and left common femoral vein for insertion of the IVUS catheter. The authors found no difference in number of passes, fluoroscopy time or rate of inadvertent needle puncture in the two groups. Recently, three reports of retrospective single center case series were published in the Journal of Vascular and Interventional Radiology (Table 1).

Table 1

The first of the three reports described a series of TIPS cases that were performed using conventional (n = 26), marker wire (n = 18) and IVUS (n = 24) guidance.17 Marker wire cases involved transhepatic or trans-splenic placement of an 0.018 inch wire into the portal vein through a 21 gauge Chiba needle using fluoroscopic and percutaneous ultrasound guidance. Cases were performed with the Colapinto needle from a right internal jugular access. There were 9 total operators, 8 of whom performed IVUS guided TIPS in addition to conventional TIPS. There were 6 direct portocaval shunt cases included, 3 performed with marker wire and 3 with IVUS guidance. Two dual TIPS cases were included, one using conventional and the other with marker wire guidance. Eleven cases involved variceal embolization which were equally distributed among the three groups. The authors found a decreased amount of fluoroscopy time, air kerma, contrast agent volume, and total procedure time when using IVUS on post hoc pairwise comparison with fluoroscopy. In fact, the use of IVUS was also associated with decreased fluoroscopy time, contrast agent volume and total procedure time even when compared to marker wire. The decreased procedure time could not be attributed to portal vein access time as the 16 minutes difference in the means did not reach statistical difference. Technical success, hemodynamic success and mean portosystemic gradient reduction were similar in the two groups. Likewise, no statistically significant difference in procedural complications was discovered. There were, however more cases of stent malpositioning and contrast agent-induced renal failure in the IVUS group even though IVUS provides an additional modality to assess stent positioning and generally decreases contrast volume usage during TIPS. These may be cases of correlation rather than causation, as IVUS likely was selected in a patient that would have been prone to renal failure or increased rate of stent malposition due to the use of less conventional routes, such as middle hepatic vein to left portal vein, that are more easily achieved with IVUS.

The second paper described a case series of 49 conventional and 40 IVUS TIPS.7 Cases were performed with the Rosch Uchida transjugular access set from a right internal jugular access. The majority of IVUS access sites were at the right common femoral vein, although two cases involved two right internal jugular access sites for the liver access set and IVUS catheter. There were 4 conventional operators and 2 IVUS operators. Direct portocaval shunts were excluded from the analysis. All cases involving tran-sjugular liver biopsy, variceal embolization, portal venoplasty and thrombolysis were excluded. The authors found statistically significant decreases in number of needle passes, contrast volume, fluoroscopy time, radiation dose and procedure time associated with IVUS. There was no difference in portosystemic gradient reduction.

The third paper described the largest case series of IVUS guided TIPS to date, with 55 IVUS and 54 conventional TIPS.21 Cases were performed using the Haskal Transjugular Portal Access set which includes a 16 gauge Ross modified Colapinto needle. Overall, there was decreased radiation dose and portal access time, defined as the time from hepatic venogram to portal venogram, associated with IVUS guidance. However, when further subset analysis was done between experienced and inexperienced operators, defined as having performed fewer than 20 TIPS as the primary operator prior to the study period, no difference was seen in the experienced operators with the use of IVUS. The authors also noted gradual decreases in portal access time over the three year study period in the IVUS group, suggestive of an initial learning curve. Further, there were fewer needle pass-related capsular perforation rates in the IVUS group.

It is important to remember that the existing literature is limited to retrospective single-center case series and findings may not be widely reproducible. These results likely reflect certain institutional factors, including practice models, operator familiarity and preferences. While the human studies listed above featured generally well-matched patient cohorts in terms of age, severity and etiology of liver disease, there were proportional differences in the indications for TIPS in several papers. Although the differences did not meet statistical significance, the distinction in TIPS indications is not insignificant, as TIPS for refractory variceal hemorrhage is more often performed on an urgent or emergent basis compared to refractory ascites. Furthermore, these case series featured varying overlap between the IVUS and conventional operators. In some papers, almost all TIPS operators performed both IVUS and conventional TIPS whereas in others, some operators exclusively performed TIPS with either IVUS guidance or conventional technique. It is not entirely clear whether the conventional and IVUS cases were performed at the same rates over study periods. There likely was a shift towards using IVUS as commonly seen with any new modality and the results may at least partially reflect study period-related confounding variables.

Conclusions

Although the current literature on the use of IVUS in TIPS is limited to single-center retrospective series, they suggest that the use of IVUS is associated with reduced portal access time, procedure time, contrast volume and radiation dosage. Available data on numbers of needle passes and rates of capsular perforation are mixed. Despite the limitations of the available reports, the authors believe that the use of IVUS should be strongly considered whenever there is known obstructing tumor mass, unfavorable vascular anatomy, Budd-Chiari syndrome and hepatic or portal venous thrombosis. Together, the recent literature provides evidence supporting the broader use of IVUS in TIPS regardless of the presence of complicating factors. The improved visualization afforded by IVUS has meant that historical relative contraindications to TIPS, such as Budd-Chiari syndrome and portal vein thrombosis are no longer effective. In fact, TIPS is used now as part of the treatment of portal venous thrombosis and/or Budd Chiari syndrome in certain cases. Accurate and comprehensive visualization of the hepatic vasculature is highly beneficial for the safety and efficacy of TIPS creation. IVUS represents an important, and some may argue crucial, modality in the TIPS interventionalist’s armamentarium.

Acknowledgments

Authors’ contributions: S.D.K. and E.W.L. equally contributed to this paper with conception and design of the study, data collection and analysis, statistical analysis, literature review and analysis, drafting and critical revision and editing, and final approval of the final version.

Conflicts of Interest

No potential conflict of interest relevant to this article was reported.

Article information

Gastrointestinal Intervention.Apr 30, 2018; 7(1): 9-13.
Published online 2018-04-30. doi:  10.18528/gii180008
Division of Interventional Radiology, Department of Radiology, UCLA Medical Center, David Geffen School of Medicine at UCLA, Los Angeles, CA, USA
Division of Interventional Radiology, Department of Radiology, UCLA Medical Center, David Geffen School of Medicine at UCLA, 757 Westwood Plaza, Suite 2125, Los Angeles, CA 90095, USA. E-mail address:EdwardLee@mednet.ucla.edu (E.W. Lee). ORCID: https://orcid.org/0000-0003-0418-1454
Received March 12, 2018; Accepted March 24, 2018.
Articles from Gastrointestinal Intervention are provided here courtesy of Gastrointestinal Intervention

References

  • Rösch J, Hanafee WN, Snow H. Transjugular portal venography and radiologic portacaval shunt: an experimental study. Radiology. 1969;92:1112-4.
  • Nishimine K, Saxon RR, Kichikawa K, Mendel-Hartvig J, Timmermans HA, Shim HJ. Improved transjugular intrahepatic portosystemic shunt patency with PT-FE-covered stent-grafts: experimental results in swine. Radiology. 1995;196:341-7.
  • Sommer CM, Gockner TL, Stampfl U, Bellemann N, Sauer P, Ganten T. Technical and clinical outcome of transjugular intrahepatic portosystemic stent shunt: bare metal stents (BMS) versus viatorr stent-grafts (VSG). Eur J Radiol. 2012;81:2273-80.
  • Smith M, Durham J. Evolving indications for tips. Tech Vasc Interv Radiol. 2016;19:36-41.
  • Miller DL, Balter S, Cole PE, Lu HT, Schueler BA, Geisinger M. Radiation doses in interventional radiology procedures: the RAD-IR study: part I: overall measures of dose. J Vasc Interv Radiol. 2003;14:711-27.
  • Kaufman JA, Lee MJ. Vascular and interventional radiology. Philadelphia: Elsevier Saunders; 2013
  • Kao SD, Morshedi MM, Narsinh KH, Kinney TB, Minocha J, Picel AC. Intra-vascular ultrasound in the creation of transhepatic portosystemic shunts reduces needle passes, radiation dose, and procedure time: a retrospective study of a single-institution experience. J Vasc Interv Radiol. 2016;27:1148-53.
  • Freedman A, Sanyal A. Complications of transjugular intrahepatic portosystemic shunts. Semin Interv Radiol. 1994;11:161-77.
  • Gaba RC, Khiatani VL, Knuttinen MG, Omene BO, Carrillo TC, Bui JT. Comprehensive review of TIPS technical complications and how to avoid them. AJR Am J Roentgenol. 2011;196:675-85.
  • Haskal ZJ, Cope C, Shlansky-Goldberg RD, Soulen MC, Baum RA, Redd DC. Transjugular intrahepatic portosystemic shunt-related arterial injuries: prospective comparison of large- and small-gauge needle systems. J Vasc Interv Radiol. 1995;6:911-5.
  • Krajina A, Hulek P, Fejfar T, Valek V. Quality improvement guidelines for transjugular intrahepatic portosystemic shunt (TIPS). Cardiovasc Intervent Radiol. 2012;35:1295-300.
  • Kably I, Pereira K, Zhong L, Cekic M. Endovascular management of hepatic arterial injury during TIPS placement. Diagn Interv Imaging. 2016;97:673-5.
  • Zhang Y, Zhang XM, Prowda JC, Zhang HL, Sant’anna Henry C, Shih G. Changes in hepatic venous morphology with cirrhosis on MRI. J Magn Reson Imaging. 2009;29:1085-92.
  • Sussman SK, Braun SD, Perlmutt LM, Newman GE, Dunnick NR. Digital indirect portography. AJR Am J Roentgenol. 1986;147:39-43.
  • Maleux G, Nevens F, Heye S, Verslype C, Marchal G. The use of carbon dioxide wedged hepatic venography to identify the portal vein: comparison with direct catheter portography with iodinated contrast medium and analysis of predictive factors influencing level of opacification. J Vasc Interv Radiol. 2006;17:1771-9.
  • Yamagami T, Tanaka O, Yoshimatsu R, Miura H, Suzuki T, Hirota T. Hepatic artery-targeting guidewire technique during transjugular intrahepatic portosystemic shunt. Br J Radiol. 2011;84:315-8.
  • Gipson MG, Smith MT, Durham JD, Brown A, Johnson T, Ray CE. Intravascular US-guided portal vein access: improved procedural metrics during TIPS creation. J Vasc Interv Radiol. 2016;27:1140-7.
  • Jongbloed MR, Schalij MJ, Zeppenfeld K, Oemrawsingh PV, van der Wall EE, Bax JJ. Clinical applications of intracardiac echocardiography in interventional procedures. Heart. 2005;91:981-90.
  • Kew J, Davies RP. Intravascular ultrasound guidance for transjugular intrahepatic portosystemic shunt procedure in a swine model. Cardiovasc Intervent Radiol. 2004;27:38-41.
  • Farsad K, Fuss C, Kolbeck KJ, Barton RE, Lakin PC, Keller FS. Transjugular intrahepatic portosystemic shunt creation using intravascular ultrasound guidance. J Vasc Interv Radiol. 2012;23:1594-602.
  • Iyamu I, Anene A, Reddick M, Anderson ME, Kalva SP, Pillai AK. Clinical utility of intravascular ultrasound guided portal vein access during transjugular intrahepatic portosystemic shunt (TIPS) creation: comparison with conventional technique. J Vasc Interv Radiol. 2015;26:S94.

Figure 1


(A) Hepatic arterial cannulation (arrowhead) and associated colonic puncture with resulting intraluminal contrast leakage (arrow) during a conventional transjugular intrahepatic portosystemic shunt procedure. (B) Follow-up computed tomography performed 6 hours following the procedure confirmed presence of contrast within bowel (arrow) and trace amounts in the peritoneum, corresponding to locations seen on intra-procedural fluoroscopy.

Figure 2


A 51-year-old female with a history of hepatitis C virus infection status post remote orthotopic liver transplant complicated by recurrent cirrhosis, esophageal varices and presenting with hematochezia and hematemesis. (A–C) Computed tomography performed prior to the procedure demonstrated large portal venous thrombus with evidence of cavernous transformation (arrowheads). (D) Fluoroscopic images demonstrating successful transjugular intrahepatic portosystemic shunt placement with the help of intravascular ultrasound (arrowhead).

Figure 3


A 57-year-old obese female with hematemesis secondary to extensive acute mesenteric, splenic and portal venous thrombosis following sleeve gastrectomy. (A, B) Computed tomography performed prior to the procedure demonstrated acute thrombosis of the portal veins (arrowheads). The presence of chronic portal venous thrombosis and comorbid conditions would have made conventional transjugular intrahepatic portosystemic shunt (TIPS) placement more difficult and risky. The decision was made to attempt portal decompression with a TIPS under intravascular ultrasound guidance. Intraprocedural intravascular sonographic images of the right portal vein (RPV) taken during evaluation (C) and needle advancement (D). The needle was advanced towards the RPV (E) to an intraluminal position (F). Digital subtraction angiograms demonstrating extensive splenic, mesenteric and portal venous thrombosis (arrowheads) following needle puncture and cannulation of the portal vein (G, H) and stent placement (I, J).

Table 1

A Summary of Outcomes of IVUS-guided Transjugular Intrahepatic Portosystemic Shunt in the Literature

First author (year) No. of IVUS No. of conventional Needle passes Portal vein access time (min) Fluoroscopy time (min) Contrast volume (mL) Procedure time (min) Air kerma (mGy) Dose area product Capsular perforation rate (%)
Farsad 20 (2012) 25 75 4/5 - 1 (30/30) - - - - 4/4
Iyamu 21 (2015) 55 54 - 0.67 (31/46) - - - 0.67 (1,592/2,376) - 9/31
Gipson 17 (2016) 24 26 - 0.74 (50/67) 0.5 (20/40) 0.63 (134/212) 0.71 (108/153) 0.4 (404/1,005) 0.64 (168/260) (Gy·cm2) -
Kao 7 (2016) 40 49 2/6 - 0.56 (19/34) 0.41 (57/140) 0.69 (86/125) 0.18 (174/981) 0.18 (3,793/21,414) (μGy·cm2) -

Values are presented as number only, number of IVUS/number of conventional, or ratio (IVUS/conventional). Actual reported values are presented in parentheses. Mean values are displayed for Farsad’s and Iyamu’s reports. Median values are displayed for Gipson’s and Kao’s reports.

IVUS, intravascular ultrasound; -, not reported.