Stents Tested and Expansion Force Averaged over Three Compression Cycles
Test-No. | Make | Model | Weave | Cover | Cover type | Diameter (mm) | Length (mm) | Force (N) |
---|---|---|---|---|---|---|---|---|
6 | Boston | Wallstent | Braided | Bare | - | 22 | 90 | 5.6 |
5 | Boston | Wallflex | Braided | Bare | - | 25 | 60 | 5.8 |
9 | Cook | Evolution | Braided | Bare | - | 25 | 100 | 7.9 |
14 | MI Tech | CNZ | Knitted | Bare | - | 24 | 60 | 8.7 |
23 | Ella | Enterella | Braided | Bare | - | 25 | 112 | 13.8 |
12 | MI Tech | TLC | Double knitted | Part | Membrane | 24 | 80 | 14.9 |
13 | MI Tech | NCS | Knitted | Bare | - | 24 | 80 | 15.4 |
17 | BCM | Hilzo bone | Knitted | Part | Membrane | 20 | 120 | 16.9 |
19 | Micro-Tech | Colon | Braided | Bare | - | 30 | 80 | 17.2 |
8 | S&G Biotech | Egis covered | Double knitted | Part | Membrane | 24 | 110 | 17.5 |
7 | S&G Biotech | Egis bare | Double knitted | Bare | - | 24 | 110 | 20.5 |
16 | BCM | Hilzo straight | Double knitted | Bare | - | 24 | 120 | 20.9 |
18 | BCM | Hilzo bone | Knitted | Fully | Membrane | 20 | 120 | 24.4 |
21 | Micro-Tech | Rectum | Braided | Bare | - | 30 | 80 | 31.8 |
15 | MI Tech | HRC (flare) | Knitted | Part | Dipped | 20 | 100 | 32.9 |
4 | TaeWoong | ComVi Flare | Double knitted | Part | Membrane | 26 | 95 | 33.3 |
2 | TaeWoong | D-stent | Knitted | Bare | - | 28 | 90 | 33.4 |
11 | MI Tech | CCBA | Braided | Fully | Dipped | 20 | 100 | 35.1 |
3 | TaeWoong | ComVi | Double knitted | Part | Membrane | 26 | 100 | 45.9 |
25 | Ella | BD | Braided | Bare | - | 26 | 60 | 53.3 |
10 | MI Tech | CCI | Braided |
Fully | Dipped | 24 | 100 | 61.7 |
24 | Ella | Enterella | Braided | Fully | Dipped | 30 | 135 | 84.4 |
1 | TaeWoong | S-stent | Braided | Bare | - | 28 | 80 | 85.1 |
22 | Micro-Tech | Rectum | Braided | Part | Dipped | 30 | 80 | 104.9 |
20 | Micro-Tech | Colon | Braided | Fully | Dipped | 30 | 80 | 130.8 |
Reference esophageal stents | ||||||||
29O | S&G Biotech | Egis | Knitted | Fully | Membrane | 20 | 110 | 39.3 |
28O | Ella | HV+ | Braided | Fully | Dipped | 18 | 110 | 133.8 |
Numbering is in order of test performed, stents are listed in ascending order of radial force.
*CCI stent classified as ‘braided’ due to mechanical behavior as a consequence of the silicone covering, despite a knitted underlying wire construction.